A theranostic 3D ultrasound imaging system for high resolution image-guided therapy

Hanna Bendjador, Josquin Foiret, Robert Wodnicki, Douglas N. Stephens, Zoe Krut, Eun Yeong Park, Zulma Gazit, Dan Gazit, Gadi Pelled, Katherine W. Ferrara*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Microbubble contrast agents are a diagnostic tool with broad clinical impact and an increasing number of indications. Many therapeutic applications have also been identified. Yet, technologies for ultrasound guidance of microbubble-mediated therapy are limited. In particular, arrays that are capable of implementing and imaging microbubble-based therapy in three dimensions in real-time are lacking. We propose a system to perform and monitor microbubble-based therapy, capable of volumetric imaging over a large field-of-view. To propel the promise of the theranostic treatment strategies forward, we have designed and tested a unique array and system for 3D ultrasound guidance of microbubble-based therapeutic protocols based on the frequency, temporal and spatial requirements. Methods: Four 256-channel plane wave scanners (Verasonics, Inc, WA, USA) were combined to control a 1024-element planar array with 1.3 and 2.5 MHz therapeutic and imaging transmissions, respectively. A transducer aperture of ∼40×15 mm was selected and Field II was applied to evaluate the point spread function. In vitro experiments were performed on commercial and custom phantoms to assess the spatial resolution, image contrast and microbubble-enhanced imaging capabilities. Results: We found that a 2D array configuration with 64 elements separated by λ-pitch in azimuth and 16 elements separated by 1.5λ-pitch in elevation ensured the required flexibility. This design, of 41.6 mm × 16 mm, thus provided both an extended field-of-view, up to 11 cm x 6 cm at 10 cm depth and steering of ±18° in azimuth and ±12° in elevation. At a depth of 16 cm, we achieved a volume imaging rate of 60 Hz, with a contrast ratio and resolution, respectively, of 19 dB, 0.8 mm at 3 cm and 20 dB and 2.1 mm at 12.5 cm. Conclusion: A single 2D array for both imaging and therapeutics, integrated with a 1024 channel scanner can guide microbubble-based therapy in volumetric regions of interest.

Original languageEnglish
Pages (from-to)4949-4964
Number of pages16
JournalTheranostics
Volume12
Issue number11
DOIs
StatePublished - 2022
Externally publishedYes

Bibliographical note

Publisher Copyright:
© 2022 Ivyspring International Publisher. All rights reserved.

Keywords

  • array design
  • CPS imaging
  • image-guided therapy
  • microbubble imaging
  • Volumetric ultrasound imaging

Fingerprint

Dive into the research topics of 'A theranostic 3D ultrasound imaging system for high resolution image-guided therapy'. Together they form a unique fingerprint.

Cite this