TY - JOUR
T1 - Accuracy of Toric Intraocular Lens Calculations Using Estimated Versus Measured Posterior Corneal Astigmatism
AU - Segers, Maartje H.M.
AU - Abulafia, Adi
AU - Webers, Valentijn S.C.
AU - Verstraaten, Jan Willem
AU - Vandevenne, Magali M.S.
AU - Berendschot, Tos T.J.M.
AU - Kan-tor, Yoav
AU - Benjamini, Yuval
AU - van den Biggelaar, Frank J.H.M.
AU - Barrett, Graham D.
AU - Nuijts, Rudy M.M.A.
AU - Dickman, Mor M.
N1 - Publisher Copyright:
© 2023 The Author(s)
PY - 2024/6
Y1 - 2024/6
N2 - Purpose: To compare the prediction accuracy of toric intraocular lens calculations using estimated vs measured posterior corneal astigmatism (PCA). Design: Retrospective case series. Methods: A total of 110 eyes of 110 patients with uncomplicated toric intraocular lens implantation were included in this study. Predicted postoperative refractive astigmatism was calculated with the Barrett Toric Calculator using the estimated PCA (E-PCA), the measured IOLMaster 700 PCA (I-PCA), and the measured Pentacam PCA (P-PCA). Refractive astigmatism prediction errors (RA-PEs), including their trimmed (tr-) centroid (mean vector), spread (precision), tr-mean absolute RA-PE (accuracy), and percentage within a certain threshold, were determined using vector analysis and compared between groups. Setting: University Eye Clinic, Maastricht University Medical Center+, the Netherlands. Results: The tr-centroid RA-PEs of the E-PCA (0.02 diopter [D] at 82.2°), the I-PCA (0.08 D at 35.5°), and the P-PCA (0.09 D at 69.1°) were significantly different from each other (P < .01), but not significantly different from zero (P = .75, P = .05, and P = .05, respectively). The E-PCA had the best precision (tr-mean 0.40 D), which was not significantly lower than the I-PCA (0.42 D, P = .53) and P-PCA (0.43 D, P = .06). The E-PCA also had the best accuracy (0.40 D), which was not significantly different from the I-PCA (0.42 D, P = .26) and significantly better than the P-PCA (0.44 D, P < .01). The precision and accuracy of the I-PCA did not significantly differ from those of the P-PCA. There were no statistically significant differences in the percentage of eyes within a certain absolute RA-PE threshold. Conclusions: The Barrett Toric Calculator using the E-PCA, I-PCA, or P-PCA showed a comparable prediction of postoperative refractive astigmatism in standard clinical practice.
AB - Purpose: To compare the prediction accuracy of toric intraocular lens calculations using estimated vs measured posterior corneal astigmatism (PCA). Design: Retrospective case series. Methods: A total of 110 eyes of 110 patients with uncomplicated toric intraocular lens implantation were included in this study. Predicted postoperative refractive astigmatism was calculated with the Barrett Toric Calculator using the estimated PCA (E-PCA), the measured IOLMaster 700 PCA (I-PCA), and the measured Pentacam PCA (P-PCA). Refractive astigmatism prediction errors (RA-PEs), including their trimmed (tr-) centroid (mean vector), spread (precision), tr-mean absolute RA-PE (accuracy), and percentage within a certain threshold, were determined using vector analysis and compared between groups. Setting: University Eye Clinic, Maastricht University Medical Center+, the Netherlands. Results: The tr-centroid RA-PEs of the E-PCA (0.02 diopter [D] at 82.2°), the I-PCA (0.08 D at 35.5°), and the P-PCA (0.09 D at 69.1°) were significantly different from each other (P < .01), but not significantly different from zero (P = .75, P = .05, and P = .05, respectively). The E-PCA had the best precision (tr-mean 0.40 D), which was not significantly lower than the I-PCA (0.42 D, P = .53) and P-PCA (0.43 D, P = .06). The E-PCA also had the best accuracy (0.40 D), which was not significantly different from the I-PCA (0.42 D, P = .26) and significantly better than the P-PCA (0.44 D, P < .01). The precision and accuracy of the I-PCA did not significantly differ from those of the P-PCA. There were no statistically significant differences in the percentage of eyes within a certain absolute RA-PE threshold. Conclusions: The Barrett Toric Calculator using the E-PCA, I-PCA, or P-PCA showed a comparable prediction of postoperative refractive astigmatism in standard clinical practice.
UR - http://www.scopus.com/inward/record.url?scp=85187317631&partnerID=8YFLogxK
U2 - 10.1016/j.ajo.2023.11.016
DO - 10.1016/j.ajo.2023.11.016
M3 - ???researchoutput.researchoutputtypes.contributiontojournal.article???
C2 - 38000619
AN - SCOPUS:85187317631
SN - 0002-9394
VL - 262
SP - 107
EP - 113
JO - American Journal of Ophthalmology
JF - American Journal of Ophthalmology
ER -