Abstract
We have developed Johnson noise thermometry applicable to mesoscopic devices with variable source impedance with high bandwidth for fast data acquisition. By implementing differential noise measurement and two-stage impedance matching, we demonstrate noise measurement in the frequency range of 120 MHz-250 MHz with a wide sample resistance range of 30 ω-100 kω tuned by gate voltages and temperature. We employed high-frequency, single-ended low noise amplifiers maintained at a constant cryogenic temperature in order to maintain the desired low noise temperature. We have achieved thermometer calibration with temperature precision up to 650 μK measuring a 200 mK temperature modulation on a 10 K background with 30 s of averaging. Using this differential noise thermometry technique, we measured thermal conductivity on a bilayer graphene sample spanning the metallic and semiconducting regimes in a wide resistance range, and we compared it to the electrical conductivity.
Original language | English |
---|---|
Article number | 014904 |
Journal | Review of Scientific Instruments |
Volume | 92 |
Issue number | 1 |
DOIs | |
State | Published - 1 Jan 2021 |
Externally published | Yes |
Bibliographical note
Publisher Copyright:© 2021 Author(s).