Prolonged 'phantom' square wave capnograph tracing after patient disconnection or extubation: Potential hazard associated with the siemens servo 900c ventilator

Yehuda Ginosar*, Dimitry Baranov

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

4 Scopus citations

Abstract

Background: The authors report on the appearance of misleading square wave 'phantom' capnograph tracings for approximately 3 min after disconnection from the Siemens Servo 900c ventilator. A series of experiments are described to examine the mechanism of this phenomenon. Methods: Patients were ventilated using the Siemens Servo 900c ventilator with the following settings: minute volume, 5 l/min; respiratory rate, 8 breaths/rain; PEEP, 0 cm H2O; trigger sensitivity, 20 cm H2O. The ventilator was connected to the Siemens Servo Evac 180 evacuation system (25 l/min on evacuation flowmeter). Airway pressure and capnography were recorded at the Y piece during ventilation and after disconnection. A back-up ventilator was used to support the patient during disconnection of the ventilator being studied. Results: Initially, the 'phantom' capnograph tracing closely resent bled the square wave capnograph tracing before disconnection, but the amplitude and shape of the waveform gradually decayed. Based on experiments described in this article, the authors show that the carbon dioxide for the 'phantom' capnograph tracing comes from the gas exhaled by the patient in the last breaths before disconnection and which is present in both the expiratory tubing and in the evacuation system. The small pressure gradient between the exhaust reservoir and the atmosphere causes reverse flow of expired gas after disconnection, when both the nonreturn flap valve at the exhaust outlet is open (due to minimal valve incompetence) and when the expiratory servo valve is open (in the absence of positive end-expiratory pressure). This continuous reverse flow is detected by the capnograph but is interrupted intermittently by each attempt positive pressure ventilation, thereby creating a 'phantom' capnograph. Conclusions: After 'accidental disconnection of the patient from the breathing system, or after accidental extubation of the trachea, the 'phantom' capnograph is likely to confuse even an experienced anesthesiologist into the mistaken belief that his rapidly deteriorating patient is being ventilated adequately. Several potential mechanism to eliminate this phenomenon are outlined, including the avoidance of zero positive end-expiratory pressure. 'Phantom' capnography provides an illustration of the dangers of using monitoring techniques, however reliable, as a substitute for vigilant clinical observation.

Original languageEnglish
Pages (from-to)729-735
Number of pages7
JournalAnesthesiology
Volume86
Issue number3
DOIs
StatePublished - Mar 1997
Externally publishedYes

Keywords

  • Carbon dioxide
  • Complications
  • Monitoring
  • Ventilation, mechanical

Fingerprint

Dive into the research topics of 'Prolonged 'phantom' square wave capnograph tracing after patient disconnection or extubation: Potential hazard associated with the siemens servo 900c ventilator'. Together they form a unique fingerprint.

Cite this