Abstract

A novel diagnostic design seeks to enhance the capability of multienergy soft x-ray (SXR) detection by using an image intensifier to amplify the signals from a larger set of filtered x-ray profiles. The increased number of profiles and simplified detection system provides a compact diagnostic device for measuring Te in addition to contributions from density and impurities. A single-energy prototype system has been implemented on NSTX, comprised of a filtered x-ray pinhole camera, which converts the x-rays to visible light using a CsI:Tl phosphor. SXR profiles have been measured in high performance plasmas at frame rates of up to 10 kHz, and comparisons to the toroidally displaced tangential multi-energy SXR have been made.

Original languageEnglish
Article number10E502
JournalReview of Scientific Instruments
Volume81
Issue number10
DOIs
StatePublished - Oct 2010
Externally publishedYes

Fingerprint

Dive into the research topics of 'Prototype high resolution multienergy soft x-ray array for NSTX'. Together they form a unique fingerprint.